2026

11th IEEE RAS/EMBS International Conference for Biomedical Robotics and Biomechatronics (BioRob)

Lab2Wear: an Open-Design Synchronization Platform for Off-the-Shelf Wearable Motion Analysis Systems

Qingya Zhao, Ton T. H. Duong, Kai-Chun Liu, Kimberly Kontson, Damiano Zanotto

Dept. of Mechanical Engineering, Stevens Inst. of Technology, Hoboken, USA

Keywords

opengo, moticon, sensor insole,time alignment,bluetooth synchronization, validation

Abstract

Wearable biosensors are increasingly utilized in clinical research to enable real-world monitoring of functional performance metrics, known as digital mobility outcomes (DMOs). As new DMOs emerge, validation tools are needed to demonstrate validity and encourage adoption. However, the lack of universal wireless-to-wired interfaces complicate synchronization with reference laboratory instruments, hindering functional validation. This paper introduces Lab2Wear, a modular, expandable, open-design platform that achieves application-layer time alignment between any off-the-shelf, API-enabled Bluetooth Low Energy (BLE) biosensor and transistor-transistor logic (TTL)-enabled laboratory instruments, without requiring firmware or hardware modifications. Lab2Wear comprises an Android Sync App and a Linux single-board computer (Sync Module), each running a dedicated Data-Driven Delay Compensation Model (D3CM) to enable wearable-to-Android and TTL-to-Android time alignment, respectively. In hour-long tests, Lab2Wear achieved TTL-to-Android MAE of 1.17 ± 0.37 ms, remained robust under increased phone-side load (1.40 ± 0.42 ms), and outperformed an Original Equipment Manufacturer (OEM) sync box (Moticon OpenGo) with end-to-end MAE of 7.26 ± 1.40 ms, within that device's sampling interval. These results support the proof-of-concept feasibility of Lab2Wear as a new open tool for independent, reproducible functional validation of API-enabled commercial biosensors.

Moticon's Summary

This validation study evaluated the Lab2Wear platform to improve timealignment for commercial biosensors. The researchers used opengo devices from moticon as a representative sensorinsole system to test end-to-end bluetoothsynchronization. By logging acceleration data from the insoles and comparing it against a mock laboratory instrument, they demonstrated that the Lab2Wear platform significantly outperformed the uncorrected OEM sync box. The findings highlight that correcting clock drift in these insoles is essential for precise data synchronization.

Contact Us
Book a free online demo or use the contact form to get in touch
Newsletter
Subscribe to our newsletter for regular updates

The form was sent successfully.

You will be contacted shortly.

You Would Like to Get in Touch?

Write us a message on product related questions or with regards to your application.  We are here to assist!


Schedule an Online Demo

Get a hands-on overview of our products, find the best choice, discuss your application and ask questions.

30 minutes

Web conferencing details will be provided upon confirmation

You are currently viewing a placeholder content from Google Calendar. To access the actual content, click the button below. Please note that doing so will share data with third-party providers.

More Information

Stay one step ahead!

Subscribe to our newsletter for the latest information on case studies, webinars, product updates and company news

Get Support

Check our FAQ database for answers to frequently asked questions

Describe your issue in as much detail as possible. Include screenshots or files if applicable.