Total Force Validation Study for Moticon SCIENCE Sensor Insole (insole3)

Schlagwörter

Zusammenfassung

Study Design

Descriptive Study

Background

Both clinicians and researchers often utilize treadmills to analyze spatiotemporal and biomechanical factors during running. However, there is question of whether or not treadmill running mimics overground running. The development of new wearable technology, such as pressure sensor insoles, presents
an opportunity to compare the two running conditions.

Purpose

To determine the validity of Moticon’s SCIENCE Sensor Insole (model: insole3) force output and its force zero level behavior during walking at normal speed. The study
quantifies the error of total force measurements compared to the output of force plates.

Methods

Ten healthy subjects (age 33.4, SD=7.68, 30% female) carried out trials to capture multiple left stance periods and the subsequent right stance periods during a longer walking sequence. The setup comprises a chair and two Hawkin Dynamics force plates (Hawkin Dynamics LLC, Westbrook, USA) which are placed in a bilateral configuration. Six runs were carried out with each subject, each 3 runs without specific calibration and 3 runs with the standard total force calibration routine included in the Moticon SCIENCE Mobile App. Zero levels of the pressure sensors were analyzed during the walking as of after a 2-minute warm up period in sitting.

Results

Compared to the force plate readings, insole3 force data of overground walking deviated 3.51% regarding average peak error (SD=4.78%) and 0.62% regarding average NMAE (normalized mean absolute error, SD=3.27%) using the standard calibration routine. No single peak error deviated more than 10.4% from the force plate reading. No NMAE error deviated more than 2.66% from the force plate reading. An average peak error of 20.2% (SD=6.11%) and a NMAE of 7.48% (SD=3.14%) was observed in the runs without the standard calibration routine. The zero levels resulted in an overall average of 5.8N for the uncalibrated runs and 6.4N for the calibrated runs.

Levels of Evidence

Inhouse study

Moticon's Schlussfolgerungen

Kontaktiere uns
Buche eine kostenlose Online-Demo oder verwende das Kontaktformular, um mit uns in Kontakt zu treten.
Newsletter
Trage Dich in unseren Newsletter ein, um regelmäßige Updates zu erhalten
Verwendete Produkte von Moticon

Select your desired system

The cutting edge test based outcome assessment system for health professionals and trainers

The most versatile toolkit for free data acquisition and comprehensive analytics in research

Have a general inquiry?

Write us a message for general questions about products and solutions or if you’d like to discuss other topics.


Wähle Deine Produktelinie

Mobile Ganganalysen und sportmotorische Tests für Trainer und Therapeuten – objektiv und effizient.

Das flexible System für freie Datenerfassung und fortgeschrittene Analysen in der Forschung.

Du möchtest eine Anfrage stellen?

Schreibe uns eine Nachricht zu allgemeinen Fragen über Produkte oder zu anwendungsbezogenen Themen, die Du besprechen möchtest.


Das Formular wurde erfolgreich gesendet.

Wir werden uns in Kürze mit dir in Verbindung setzen.

moticon-rego-sensor-insole-live-event

Immer einen Schritt voraus!

Abonniere unseren Newsletter für die neuesten Informationen zu Fallstudien, Webinaren, Produkt-Updates und Neuigkeiten bei Moticon

Hole dir Unterstützung

Finde Antworten zu den häufigsten Fragen in unseren FAQ

Beschreibe Dein Problem so detailliert wie möglich. Hänge Screenshots oder Daten an, sofern das hilfreich ist.


Need help?
Want a live demo?
Interested in prices?
Want to say hello?
Always just a call away
+49 89 2000 301 60
Du brauchst Unterstützung?
Du möchtest eine Live Demo?
Interessiert an Preisen?
Du hast Fragen?
Nur einen Anruf entfernt
+49 89 2000 301 60

The form was sent successfully.

You will be contacted shortly.

Get support

Check our FAQ database for answers to frequently asked questions

Describe your issue in as much detail as possible. Include screenshots or files if applicable.